Filter
The item has been added to your inquiry list
torque ratchet,
A socket for the different tools, such as screwdrivers and tighteners, is located on the head of the ratchet. It restricts the torque previously set on the tensioning screw - between 0 and 30Ncm - by slipping off the ratchet head. This torque limit can be disabled by twisting the head, in order to tighten implants into the perfect position, for example. The direction of rotation is marked on the ratchet head. We recommend taking the ratchet apart for purposes of cleaning and sterilisation.
Locator instrument, three-part,
Screwed stainless steel tip for removing retention inserts, mid section for repositioning the retention inserts, and gold coloured insertion aid for final screwing of the Locator®.
Guide key for insertion aid,
Any loosening movement of the implant is prevented by pushing up the outer hexagon on the insertion post when releasing the retaining screw. Furthermore, the insertion post can be loosened if it ever jams.
Enlargement drill 2.5 mm,
Variante:
Erweiterungsbohrer 2,5 mm
Use this extension drill without internal cooling for the preparation of implant sites for IMPLA implant lines Dual Surface and Micro Retention. The suitable drill diameter is selected from the drill protocol depending on the implant diameter and the drilling depth.The depth inside the bone is controlled during the drilling process with help of the depth marks on the drill. The drills with the diameters 3.15 mm and 4.25 mm are available in a short version for use in the molar region.
Enlargement drill 4.90 mm for Cylindrical,
Variante:
Erweiterungsbohrer 4,9 mm für Cylindrical
Use this extension drill without internal cooling for the preparation of the implant site for IMPLA Cylindrical implants. The suitable drill diameter is selected from the drill protocol depending on implant diameter and drill depth.The drill depth in the bone is controlled with help of the depth marks on the drill. The drills with diameters 3.15 mm and 4.25 mm are available in a short version for use in the molar region.
Enlargement drill short 2.5 mm,
Variante:
Erweiterungsbohrer kurz 2,5 mm
Extension drill, short, without internal cooling, diameters 2.5 mm, 3.15 mm and 4.25 mm for all IMPLA Micro Retention implant lines.Depth stops are available for all pilot drills and extension drill. These depth stops give the user security and control, even in difficult anatomical situations.
Enlargement drill short 2.75 mm (for Cylindrical),
Variante:
Erweiterungsbohrer kurz 2,75 mm für Cylindrical
Extension drill short, without internal cooling, diameters 2.75 mm, 3.0 mm, 3.7 mm, 3.9 mm, 4.6 mm and 4.9 mm for all IMPLA Cylindrical implant lines. Depth stops are available for all pilot drills and extension drills. These give the user security and control, even in difficult anatomical situations.
Countersink cutter 3.3 mm,
Variante:
Bettfräser 3,3 mm
Site drill without internal cooling for shaping the implant silhouette in the bone of IMPLA implants Micro Retention or Dual Surface.
Countersink cutter 5.25 mm for Cylindrical,
Variante:
Bettfräser 5,25 mm für Cylindrical
Gently prepare the implant site in bones of medium to high hardness. Site drill without internal cooling for shaping the implant silhouette in the bone for IMPLA Cylindrical implants.
Cortical drill 2.0/3.15 mm,
Variante:
Kortikal-Bohrer 2,0 / 3,15 mm
Without internal cooling. Ideally use on D1 and D2 bones. Cortical drill for the extension of 2 mm pilot drillings to 3.15 mm diameter of the next drill setting in the corticalis.
Pilot drill short 2.0 mm,
Total length: 33.5 mmDiameter: Ø 2.0 mm The IMPLA pilot drill will help you determine the planned implant length and insertion vector. You will find the according depth marks on the spiral drills which can be verified with the length mark keys of the surgery tray.The drill collar of the IMPLA pilot drill has been constructed to accommodate the IMPLA depth stops as an addition. The IMPLA depth stops will provide a mechanical depth stop when drilling into the jaw bone for an even higher level of safety in implantology. The drill collar of the IMPLA pilot drill as well as the according depth stop are color-coded. This facilitates a clear and easy handling.
Pilot drill 2.0 mm,
Surgical pilot drill without internal cooling for different drill settings during the preparation of the implant site for the IMPLA implant lines Cylindrical, Dual Surface and Micro Retention (according to the drill protocol, limited use for Mini and Interim).
Tissue cutter 5.3 mm,
Variante:
5,3 mm
Available in diameter 3.30, 4.20 and 5.30 mm, and perfectly aligned with the geometry of the Impla range. It has a scalpel-like, circular cutting edge, and is used for carrying out flapless insertions or extremely precise re-opening of the mucous membrane after the healing phase.
Instrument contra-angle,
Contra angle-driven 3 mm Allen key for insertion post of IMPLA implants for machine insertion. The key has a snap-in function for implants.
Drill extension,
Guarantees comfortable and safe work evenin situations with limited interdental space. For the extension of the tool shaft for contra-angle 2.34 mm.
bone thread cutter 3.3 mm/11.5 mm,
Variante:
3,3 mm / 11,5 mm
Precutting the thread reduces the screw-in torque especially in D1 and D2 bone qualities, to avoid overheating and extreme compression. Use the thread cutter considering the drilling protocol for Micro Retention and Dual Surface implants.
pilot drill 1.8 mm,
The pilot hole created by the pilot drill in the bone guides the tip of the pre-drill which follows, and keeps it from slipping in difficult to reach areas or in unfavorably shaped jaw ridges.
hex screwdriver short, width 1.2,
Variante:
1,2 mm / kurz
The required torque for the retaining screw is exerted with the help of the torque ratchet. This long version of the screwdriver can be used in jaws with teeth, in the anterior area. It has a 1.2 mm hexagon at the tip. A drilled hole for a safety cord is located in the turntable of the screwdriver grip.
screwdriver width 2.3 mm, machine-driven
The screwdriver with contra-angle connection offers a 2.3 mm hex connector. With help of the contra-angle, it is used to screw in two-piece IMPLA implants automatically and directly. The screwdriver 2.3 mm fits exactly into the inner hex of the IMPLA Hex Connection implants to insert them. Furthermore, it serves as a screw-in instrument for SAE cone adaptors.
Insertion aid for Mini implant balltop,
The tool’s working tip fits over the Torx connector of the Mini Balltop implant. It can be used by hand, or with the torque ratchet. A drilled hole safety cord is located in the turntable of the grip.
hex screwdriver short, width 2.3,
Variante:
kurz
The tool’s working tip grabs directly onto the hexagonal interior of the implant. It can be used by hand, or with the torque ratchet. The short version is used for jaws without teeth, or in cases where the mouth opening is limited. A drilled hole for a safety cord is located in the turntable of the screwdriver grip.
adapter ratchet for Locator adapter contra-angle,
Tool for manual unscrewing the Locator abutment in conjunction with 636075 in the torque ratchet.
key for insertion aid (for ratchet) short, width 3,
Variante:
Schlüssel für Eindrehhilfe kurz, SW 3
The tool’s working tip fits onto the external hexagon of the insertion post. It can be used by hand, or with the torque ratchet. The short version is used for jaws without teeth, or in cases where the mouth opening is limited. A drilled hole safety cord is located in the turntable of the grip.
Insertion aid 2.5 mm short, for Cone Connection 5.3 mm
Variante:
Eindrehhilfe 2,5 mm kurz
Ratchet tool to change the insertion depth or the hex position of IMPLA Cone Connection implants after removal of the insertion post (part of the packaging).The tip of this tool fits precisely into the implant geometry. It can be used manually or with a torque ratchet,.
Insertion aid 2.0 mm long, for Cone Connection 3.3/4.2 mm
Variante:
Eindrehhilfe 2,0 mm lang
Ratchet tool to change the insertion depth or the hex position of IMPLA Cone Connection implants after removal of the insertion post (part of the packaging).The tip of this tool fits precisely into the implant geometry. It can be used manually or with a torque ratchet,.
Depth gauge 6.5 mm - 14.5 mm,
For checking the plant implant insertion depth after use of the pilot drill 2 mm and of the extension drill according to the implant diameter. For checking the orientation / divergence of the implants.
Insertion aid machine-driven with contra-angle connector, for Mini balltop
The tool’s working tip fits over the Torx connector of the Impla Mini Balltop implant. It can be used either manually, (adapter for elbow instruments 636077) or with the contra angle.
contra-angle adapter width 1.2 mm,
Variante:
1,2 mm / universal
This can be used either manually (adapter for elbow instruments 636077) or with the elbow attachment. The short version is used for jaws without teeth, or in cases where the mouth opening is limited. It has a 1.2 mm hexagon at the tip.
Insertion aid long machine-driven 2.5 mm, (for Cone Connection 5.3 mm)
Variante:
Eindrehhilfe lang, masch. 2,5 mm
Screwdriver with connection geometry 2 mm for all Cone Connection implants. For making fine adjustments to the implant in the bone after the insertion post which comes with the implant packaging has already been removed (Cone Connection series, diameter 3.3 mm and 4.2 mm). The screwdriver can be used with a contra-angle, manually with the adapter art. no. 636077 or with the torque ratchet.
open-end pick-up,
This open ended wrench is needed for transfer of the implant into the IPK wrench, and insertion via the drilling template. The open ended wrench can also be used for non-contact removal of the implant from the packing, even where the opening of the patient’s mouth is limited (Pick Up technique).
Depth gauge 4.2mm red,
Variante:
4,2 mm
For checking the plant implant insertion depth after use of the pilot drill.